| Industrial
Process Electrodes pH
Differential Electrode
Designed for continuous
use in industrial process applications, the ECFP21A1A1 pH Differential
Electrode offers several advantages for demanding process applications:
-
Longer
sensor life in a wide range of measurement environments.
-
The reference electrode solution is buffered and less susceptible
to contamination than the standard KCL reference solutions used
in combination electrode systems.
-
Any buffer standard from 2 pH to 10 pH can be used as a reference.
This is very useful for applications that require adjustable isopotential
points.
-
The reference electrode is a pH glass bulb and the inner fill
and silver wire are not
easily contaminated. For standard combination electrode systems
silver/silver
chloride wires are used as a reference and can react with some
process chemicals
over time.
-
Reference solutions and salt bridges can be easily replaced when
fouling does
occur.
-
Integral electronic amplifiers measure potential difference of
the process electrode
and the reference electrode with respect to solution ground. This
provides a low
impedance output from the sensor which results in the following
benefits:
-
Extended cable lengths of up to several hundred feet
-
Low noise and expensive coaxial cable is not needed.
-
Hand capacitance is dramatically reduced – readings
do not fluctuate when
sensors are held.
-
Electrical interference from motors, pumps and transformers
are reduced.
-
Ground currents are essentially eliminated.
Find
out more about pH Differential Measurements

Applications
Differential
pH electrodes need special input electronics and cannot be connected
to conventional pH transmitters and controllers. A good example of
a differential pH system is the Alpha
pH 2000D Controller and the Differential pH Electrode.
Specifications
of pH
Differential Electrode
| Product
Code |
ECFP21A1A1 |
| Wetted
Materials |
Sensor
Body - CPVC
Salt Bridge Junctions - Kynar®
pH Electrode - General glass
Ground Rod - 316 Stainless Steel
O-Ring Seals - Viton®
(Consult
factory for customized material construction)
|
| Operating
Temperature |
-5°C
to 95°C (23°F to 203°F) |
| Maximum
Pressure |
100
psi @ 100°C (100 psi @ 212°F) |
| Maximum
Flow Rate |
3
m/sec (10 ft/sec) |
| Measuring
Range |
0
to 14 pH |
| Resolution |
0.01
pH |
| Standard
Sensor Cable Length |
3
m (10 ft) |
| Cable
Wire Colors |
White
Wire (-5V)
Black Wire (Ground)
Red Wire (pH Electrode/Active(Measure)
Green Wire (Reference Electrode)
Yellow Wire (Temperature/NTC 300)
Drain/Shield (Earth) |
| Maximum
Transmission Distance |
914
m (3,000 ft) |

 
Thermo
Fisher Scientific Inc.