| PRODUCT
CODE : EC-PHCTP2000D |
| (a)
pH |
| pH Measuring Range |
-2.00 to 16.00 pH |
| Resolution |
0.01 pH |
| Accuracy |
± 0.01 pH |
| (b)
mV |
| mV Measuring Range |
-1000 to 1000 mV |
| Resolution |
1 mV |
| Accuracy |
± 1 mV |
| (b)
mV percentage |
| Percentage mV Measuring Range |
0 to 100% (for -100 to 100 mV) |
| Resolution |
0.1% |
| Accuracy |
± 0.2 % (or ± 1 mV) |
| (c)
Temperature |
| Temperature measuring range |
-10.0 to +125.0 oC or +14.0 to
+257.0 oF for pt100/pt1000-10.0 to +110.0
oC or +14.0 to +230.0 oF for NTC300 |
| Resolution |
0.1 oC or 0.1 oF |
| Relative accuracy |
0.5 oC or 1.0 oF |
| Sensor |
NTC 300, Pt100 or Pt1000 (Jumper Selectable)
– 2 wire or
3 wire |
| Compensation |
Automatic (± 10 ˚C/ ± 18 ˚F offset
adjustment) or Manual |
| |
|
| Set-point
and control function – Relay 1 & Relay 2 |
| Function Switchable- Selection
in Control (CNTR) function of Setup |
OffLimit ControlP/PI
Control (Pulse length / Pulse frequency) |
| (a)
Off |
| The control function of the controller is
off |
SP1 & SP2 function will not appear in
setup when Off is selected in CNTR function to off mode |
| (b)
Limit Control & P/PI Control |
| Relay contact selection for active mode |
Energise (EN) orDe-energise (DEEN) |
| Set Point 1 (SP1) |
-2.00 to 16.00 pH or 0 to 100% or -1000 to
1000mV |
| Set Point 2 (SP2) |
-2.00 to 16.00 pH or 0 to 100% or -1000 to
1000mV |
| Active mode selection (for SP1 & SP2) |
LO (Relay is active when measured value <
set point)HI (Relay is active when measured value >
set point) |
| Hysteresis
(for SP1 & SP2) |
0.1 to 1 pH or 1 to 10.0%/ 10 to 100 mV |
| On (pickup) Delay (for SP1 & SP2) |
0 to 2000 seconds |
| Off (Dropout) Delay (for SP1 & SP2) |
0 to 2000 seconds |
| (c)
P/PI Control only |
| Proportional
Range |
10 to 500 % |
| Integral Action Time (IAT) |
0.0 to 999.9 minutes |
| (c)
(1) Pulse Length Control (PLC) |
| Pulse Length |
0.5 to 20.0 seconds |
| (c)
(2) Pulse Frequency Control (PFC) |
| Pulse Frequency |
60 to 120 pulses / minutes |
| Contact Outputs |
Two SPST relays |
| Switching Voltage / Current / Power |
Max 250VAC / Max 3A
/ Max 600VA |
| |
| Wash Function
– Wash Relay |
| Relay contact selection for active mode |
Energise (EN) orDe-energise (DEEN) |
| Wash Cycle Interval |
0.1 to 200.0 hours |
| Wash Duration |
1 to 2000 seconds |
| Contact Outputs |
One SPST relay |
| Switching Voltage / Current / Power |
Max 250VAC / Max 3A
/ Max 600VA |
| |
|
Alarm Function – Alarm Relay |
| Alarm Delay |
0 to 2000 seconds |
| Alarm control |
Steady or
leet (pulse) |
| Contact Outputs |
One SPDT relay |
| Switching Voltage / Current / Power |
Max 250VAC / Max 3A
/ Max 600VA |
| |
|
|
Transmitter Function |
| (a) Current Output 1 – pH/mV |
Proportional to pH/mV measured |
| Current output selection |
4 – 20mA or 0 – 20mA |
| pH/mV value for 0 or 4mA current |
-2.00 to 16.00 pH or 0 to 100% or -1000 to
1000mV |
| pH/mV value for 20mA current |
-2.00 to 16.00 pH or 0 to 100% or -1000 to
1000mV |
| CU22 function |
On
- 22mA current output for over range or under range or if outside
zoom settingOff
- 0 or 4 mA current output for under range
or below lower zoom setting - 20mA current output for
over range or above upper zoom setting |
| (b) Current Output 2 - Temperature |
Proportional to Temperature measured |
| Current output selection |
4 – 20mA or 0 – 20mA |
| Temperature value for 0 or 4mA current |
-10.0 to +115.0oC or +14.0
to +237.0oF for pt100/pt1000-10.0 to +100.0oC
or +14.0
to +212.0oF for NTC300 |
| Temperature value for 20mA current |
0.0 to +125.0oC or +32.0
to +257.0oF for pt100/pt10000.0 to +110.0oC
or +32.0
to +230.0oF for NTC300 |
| |
|
| Voltage
Output & Hold Function Switch |
| +12 V Output |
11.5 to 12.5 V Output - Max 50 mA |
| Hold Function Switch |
Hold function to freeze output current (0/4...20mA)
and release control relays |
| |
|
| Display |
| LCD |
UV coat, backlit 14 segments display with
symbols for status information |
| Back light |
On/Off selectable with four levels of brightness
control |
| |
|
|
Power Supply |
| Input |
80 to 250VAC/DC, 50/60Hz, Approximately
10VA |
| Main Fuse |
315mA Time Delay, 250V, Bussmann BK/GDC-315mA |
| Pollution Degree |
2 |
| Transient Voltage Category |
II |
| Connector |
5-pin screw terminal |
| |
|
|
Electrical data and Connections |
| pH/mV Input |
8-pin Screw Terminal (3.5mm pitch) |
| Temperature Input |
8-pin Screw Terminal (3.5mm pitch) |
| Relay 1, Relay 2, Wash
Relay & Alarm Relay |
9-pin Screw Terminal (7.62mm pitch) |
| Current Output 1, Current Output 2, 12V output
& Hold function switch |
13-pin Screw Terminal (3.5mm pitch) |
| Connection terminal |
5-pin, 8-pin, 9-pin & 13-pin terminal
connectors |
| |
|
|
Mechanical Specifications |
|
| Dimensions (W x H x D) - Wall
Mount |
144 x 144 x 111.5 mm |
| Weight -
Wall Mount |
950g (unit) / 1100g
(Packed) |
| Environmental Rating - Wall Mount |
NEMA 4X, IP65 |
| |
|
|
Electromagnetic Compliance (EMC) Specifications |
| Emitted Interference |
EN 61326 |
| Immunity to Interference |
EN 61326 |
| |
|
| Environmental
Conditions |
| Ambient Temperature operating range |
0 to 40 oC |
| Maximum Relative Humidity |
80% up to 31 oC decreasing linearly
to 50% at 40oC |